Rmb: 1100 3600 特惠 3600
产品规格
Safety datasheet
概述
产品名称
Human CD127/IL-7 R alpha enzyme-linked immunoassay kit
产品类型
Elisa Kit
灵敏度
10.55 pg/ml (10 μl)
| pg/ml | O.D. | Average | Corrected | |
|---|---|---|---|---|
| 0.00 | 0.0640 | 0.0630 | 0.0635 | |
| 156.25 | 0.1851 | 0.1668 | 0.1760 | 0.1125 |
| 312.50 | 0.4100 | 0.3861 | 0.3981 | 0.3346 |
| 625.00 | 0.7055 | 0.6823 | 0.6939 | 0.6304 |
| 1250.00 | 1.2610 | 1.2630 | 1.2620 | 1.1985 |
| 2500.00 | 2.0830 | 2.0410 | 2.0620 | 1.9985 |
| 5000.00 | 3.1730 | 3.1000 | 3.1365 | 3.0730 |
| 10000.00 | 4.1100 | 4.0396 | 4.0748 | 4.0113 |
精度
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average (pg/ml) | 196.2 | 1212.5 | 3366.8 | 200.9 | 1152.1 | 3404.0 |
| Standard Deviation | 6.7 | 40.4 | 193.6 | 6.5 | 57.4 | 147.8 |
| Coefficient of Variation (%) | 3.4 | 3.3 | 5.8 | 3.2 | 5.0 | 4.3 |
板内精密度:3 个已知浓度的样本酶标板内重复测定 22 次,评估酶标板内的精密度。
板间精密度:3 个已知浓度的样本酶标板间重复检测 6 次,评估酶标板间的精密度。
加标回收
健康人血清加入3个不同浓度水平的人CD127/IL-7 R alpha,未加入人CD127/IL-7 R alpha的血清作为本底,计算回收率。
回收率的范围从74%至107%,平均回收率为93%。
样本值
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | 4551.82-11441.82 | 100 | 7455.88 |
应用本试剂盒,检测30份健康人的血清样本。
n.d. =测不到浓度值。样本的浓度值低于灵敏度被认为测不到浓度值。
实验类型
Sandwich Elisa
种属反应性
Human
检测方法
Colorimetric
操作时长
120min
标准曲线梯度
7点
孵育次数
2
可检测样本
可溶性物质的液相样品。例如:血清,血浆,细胞培养上清,组织研磨液等。
加样体积
10 μl
产品特性
储存条件
存储于 2-8 ℃
靶点
UNIPROT
功能
白介素-7是一种参与调控淋巴生成的糖蛋白。细胞对IL7的反应依赖于白介素7受体(IL7R)的存在;活性受体是α/γ链异二聚体。γ(c)链也与白介素-2受体结合,主要激活IL7R复合物的信号转导,而IL7R的α链通过与细胞质信号分子结合决定特定的信号事件。人类和老鼠的串行几乎一模一样。在人类中,严重的联合免疫缺陷是由IL-7受体的遗传缺陷引起的。449位的酪氨酸残基是IL-7受体细胞内结构域的关键信号传导位点。IL-7受体被激活后,该位点被janus激酶迅速磷酸化。
同靶点&同通路的产品