Human Adiponectin enzyme-linked immunoassay kit
概述
产品名称
Human Adiponectin enzyme-linked immunoassay kit
产品类型
Elisa Kit
灵敏度
0.15 pg/ml (50 μl);3.56 pg/ml (10 μl)

pg/ml | O.D. | Average | Corrected | |
---|---|---|---|---|
0.00 | 0.0097 | 0.0102 | 0.0100 | |
6.86 | 0.0277 | 0.0283 | 0.0280 | 0.0181 |
20.58 | 0.0665 | 0.0592 | 0.0629 | 0.0529 |
61.73 | 0.1897 | 0.1883 | 0.1890 | 0.1791 |
185.19 | 0.5615 | 0.5451 | 0.5533 | 0.5434 |
555.56 | 1.4160 | 1.4190 | 1.4175 | 1.4076 |
1666.67 | 2.8940 | 2.7930 | 2.8435 | 2.8336 |
5000.00 | 4.1026 | 4.1065 | 4.1046 | 4.0946 |
精度
Intra-assay Precision | Inter-assay Precision | |||||
Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
22 | 22 | 22 | 6 | 6 | 6 | |
Average(pg/ml) | 103.3 | 485.3 | 1632.6 | 111.5 | 520.8 | 1724.6 |
Standard Deviation | 2.8 | 27.8 | 81.6 | 5.8 | 31.7 | 128.9 |
Coefficient of Variation (%) | 2.7 | 5.7 | 5.0 | 5.2 | 6.1 | 7.5 |
板内精密度:3 个已知浓度的样本酶标板内重复测定 20 次,评估酶标板内的精密度。
板间精密度:3 个已知浓度的样本酶标板间重复检测 6 次,评估酶标板间的精密度。
加标回收
健康人血清加入 3 个不同浓度水平的人Adiponectin,未加人 Adiponectin的血清作为本底,计算回收率。回收率的范围从 81 % 至 126 %,平均回收率为 93 %。
样本值
Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
---|---|---|---|---|
Serum | 30 | 2621.9-7072.8 | 100 | 4559.1 |
应用本试剂盒,检测 30 个健康志愿者的血清样本,志愿者的用药史不详。n.d. = 测不到浓度值。样本的浓度值低于灵敏度被认为测不到浓度值。
实验类型
Sandwich Elisa
种属反应性
Human
检测方法
Colorimetric
操作时长
120min
标准曲线梯度
7点/3倍
孵育次数
2
加样体积
50 μl/10 μl
产品特性
储存条件
存储于 2-8 ℃
靶点
功能
脂联素(也称为 30 kDa 脂肪细胞补体相关蛋白,Acrp30)是一种脂肪细胞来源的蛋白质,对代谢和炎症具有广泛的旁分泌和内分泌作用。它促进脂肪细胞分化、脂肪酸分解代谢和胰岛素敏感性,与肥胖、2 型糖尿病和动脉粥样硬化形成呈负相关。脂联素结合成 70 kDa 非共价连接的三聚体,这些三聚体通过二硫键结合成中等分子量六聚体,然后结合成 >300 kDa 高分子量低聚体。